The Basics Of A pH Sensor
Electrode for Measuring
The pH of an aqueous solution can be determined with the use of the measurement electrode, which is developed for this purpose.
The platinum/hydrogen pH sensor electrode has been used to measure the concentration of hydrogen ions in aqueous solutions since its invention in 1897. In addition to performing this function, it serves as a reference standard for the electrometric measurement of pH. The hydrogen electrode of a pH sensor consists of a platinized platinum plate or rod that is coated with platinum black and passed through a flow of hydrogen gas. A silver wire that has been coated with silver chloride is the reference electrode. The fundamental concept underlying the use of a pH sensor hydrogen electrode is as follows: The atoms on the surface of a metal rod can become ionized when the rod is submerged in an aqueous solution containing its own salt. A silver electrode can get ionized, for instance, when bathed in silver nitrate. The positively charged metal ions on the surface of the rod will be attracted to the negatively charged water molecules, leaving the metal negatively charged. This exchange of charges at the phase boundary between metal and solution generates a potential difference. The potential, also known as the galvanic potential, varies according to the concentration of ions in a solution. Even in modern times, the hydrogen electrode is employed as a reference standard, due in large part to the incredibly exact findings of its tests. However, for a variety of pragmatic reasons, the hydrogen electrode has lost its significance due to the fact that it is difficult and complicated to manipulate. Only the antimony electrode has lasted among all the other metal electrodes. Hydrofluoric acid, which is frequently employed to etch glass, has no effect on antimony. Precision and measuring range is nonetheless significantly constrained. Antimony, a chemical that can cause cancer in humans, must be treated with care.
The Electrode Consisting of Glass It wasn’t until the advent of the glass electrode that pH measuring became a straightforward and reliable technique that could be applied in a range of situations. In recent years, glass electrodes have exceeded all other types of indicator electrodes for pH sensor measurements, making them the predominant type of electrode used for these tests. The pH sensor measurement of an aqueous solution is now as routine as the measurement of temperature and pressure due to the dependability and accuracy of the glass electrode in conjunction with highly stable electronic amplification. This tutorial will provide you with the knowledge you need regarding the performance and maintenance of the glass electrode in order to correctly apply it. The shaft of a glass electrode is composed of glass, which must have a high resistance to hot alkaline solutions and an electrical resistance that is many times greater than that of membrane glass. The shaft of a glass electrode is composed of glass material. The hemispherically shaped electrode tip, also known as the glass membrane, is the pH-sensitive component of the glass electrode.
The membrane is made from a special glass that is sensitive to hydrogen ions, and it adheres to the shaft of the electrode. The glass electrode is partially saturated with pH Cable, a buffer solution with an average pH value of 7.
This buffer is supplemented with a predetermined amount of potassium chloride, or KCl. As a conducting electrode, a silver wire coated with silver chloride (Ag/AgCl) is inserted through the glass electrode until it reaches the internal buffer. Thus, an electrochemical circuit is formed. Using the core of the coaxial pH cable, a connection has been created between the Ag/AgCl wire and one of the terminals on a pH sensor meter. Glass Membrane Membrane All types of glass generate a potential difference proportional to the concentration of hydrogen ions in aqueous solutions. However, only particular kinds, such as the standard McInnes glass (Corning 015), can provide galvanic potentials that are consistent with the NERNST equation over a large pH range.
When the membrane glass of a measuring electrode comes into contact with an aqueous solution, a 10-4 m-thick gel layer forms between the glass surface and the solution. This gel layer prevents the solution from permeating the glass by acting as a barrier. The thickness of the gel layer is determined by the quality and composition of the membrane glass, in addition to the temperature and pH value of the solution being tested. As a result of the internal side of the glass membrane coming into contact with the inner buffer, which is an aqueous solution with a pH of 7, a gel layer forms on the interior of the membrane. On both sides of the membrane, an ongoing exchange of H+ ions occurs between the gel layers and the H+ ions present in the fluids. The outcome of this ion exchange is determined by the number of H+ ions present in one or both of the solutions. If the concentration of hydrogen ions in each solution is identical on both sides of the glass membrane, ion exchange will halt once equilibrium is established between the H+ ions in the solutions and the H+ ions in the gel layers. This holds true even if the concentration of hydrogen ions in each solution on either side of the glass barrier is the same. Consequently, both sides of the membrane glass have the same potential, and there is no difference between the potentials of the two sides.
If the concentration of hydrogen ions differs between the inner buffer and outer solution, there will be a potential differential between the inner and outer sides of the membrane glass. This potential difference’s magnitude will be proportional to the pH difference between the inner buffer and the outside solution. In order to measure the membrane’s potential, it is necessary for the membrane itself to possess a certain degree of conductivity. This is made feasible by the mobility of the alkaline ions in the membrane glass (Li+ ions in the majority of modern membrane glasses, as opposed to Na+ ions in earlier membrane glasses). Both the reaction time and the characteristic slope of the glass electrode are impacted by the thickness and composition of the gel layer. Consequently, the gel layer is of the utmost importance for the overall performance of the electrode.
A pH sensor measurement is impossible without the presence of the gel layer. Unfortunately, a gel layer requires between one and two days to grow completely. Before it can be utilized, a measuring electrode must be hydrated (immersed in ordinary, clean tap water) for at least twenty-four hours. The majority of manufacturers ship their electrodes with the membrane already hydrated (the membrane is kept moist with a KCl solution contained under a plastic cap), making the electrode immediately usable.